Colorimeters
|
|
Colorimeters |
|
COLORWAVE
CO7000 |
|

|
The
Colorwave CO7000 Medical Colorimeter is designed
for use by Doctors and medical Technologists in
small and medium sized clinics. It will withstand
high temperatures and humidity. |
|
Simple
to use:
Operation
is easy with only three buttons: reference, test
and on/off.
The
wavelength is changed by using a thumbwheel.
Designed
for clinical applications:
The
CO7000 contains 10 sealed glass filters covering
the wavelengths 400 t0 700nm making it ideal for
clinical/medical applications.
Versatile:
The
CO7000 is an 'open' system that accommodates both
conventional wet reagents and kits.
The
CO7000 requires only a small volume of reagent
(min 0.4ml) in a cuvette or tube. It can be
powered by an external power adaptor or an
internal battery making instrument portable and
usable where main electricity are unreliable. |

Click
to view larger image |
|

A
potable, battery operated unit you can take
anywhere, ideal for field clinics. |
|
COLORWAVE
CO7000 SPECIFICATIONS |
|
Wavelength
range 400
- 700nm
Filters
10
gelatin filters mounted in glass - 400, 440, 470,
490,
520, 550, 580, 590, 680, 700mm (or
optional interference filters)
Bandwidth
Typically
40nm for gelatin filters
Range
Absorbance
–0.3A to 1.99A
Accuracy
<±0.05A
at 1A
Repeatability
±0.02A
at 1A using cuvettes
Controls
3
button - on/off, Reference, Test
Cuvette
holder Fixed
with drain hole. Accepts 10mm path length
cuvettes
or 16mm round tubes. Can accept 10 & 12
mm tubes with the adaptors supplied
Power
requirements External
power adaptor (110 to 220V) or internal rechargeable
NiMH battery
Humidity/heat
protection Yes.
PCB coated. Protected to 45°C and 70% humidity
Light
source Tungsten
lamp
Approximate
dimensions 180
x 150 x 60mm
Weight
0.6kg |
|
COLORWAVE
CO7500 |
|

|
The
colourwave CO7500 is designed with students in mind.
The CO7500 is the natural successor to the well
established CO70 unit. The choice of absorbance or %
Transmission readout allows explanation of the fundamentals
of colorimetry. The integral kinetics mode permits
the study of rates of reaction and results can be
output to PC, chart recorder or printer. |
|
Simple
to use:
An
uncluttered digital display, coupled with simple
push button control make the colourwave ideal for
beginners. reading can be displayed as absorbance
Units or % transmission with one button push.
Robust:
The
colourwave CO7500 is compact and robust to
withstand rigors of the teaching environment. Eight
filters covering the range 440 - 700nm are held in
an integral filter wheel preventing any accidental
loss or damage. They can be exchanged if required.
Versatile:
When
the colourwave is switched to kinetic mode,
reading are taken approximately every second,
recording results on the display and also via the
analogue and digital outputs.
The
colourwave is available in mains and battery /
mains powered versions. |

Click
to view larger image |
|

Designed
with students in mind |
|
COLORWAVE
CO7500 SPECIFICATIONS |
|
Wavelength
range 440 -
680nm.
Standard
gelatin filters 440,
470, 490, 520, 550, 580, 590, 680nm
Bandwidth
40nm
Range:
Absorbance %
Transmission -0.3A
to 1.99A 0-199%T
Accuracy
<±0.05
at 1A
Repeatability
±0.02 at
1A using cuvettes
Operational
Modes Absorbance
Transmission Kinetics
Controls
5 button
– on/off, REF, TEST, Kinetics, Abs/%T
Cuvette
holder Fixed
with a drain hole Accepts
10mm path length square (approx.
12.5 x 12.5 overall) or 16mm round
tubes Can
accepts 10 and 12 mm tubes with adaptors
(optional)
Output
0-2V for
0-2A } via 2 x 4mm sockets }
~100mV offset in 0-1.99V
for 0-199%T} the output voltage RS232
Power
requirements 110-250V
50/60Hz 20VA via IEC connector
CO7500B only:
Rechargeable NiMH battery
Approximate
dimensions 180
x 150 x 60mm
Weight
0.5kg
(CO7500B 0.6kg) |
|
LIGHTWAVE
S2000 |
|

|
The
lightwave is uniquely shaped and perfectly tuned to
meet all your routine UV / Vis spectroscopy requirements. By
forcing on the essentials, the Lightwave gives you
more performance, more reliability, more convenience
and more productivity at a very attractive price. |
|
-
Simple programming and control with capability to
store up to 99 methods.
-
Simultaneous measurements at two wavelengths for
rapid, easy DNA / RNA analysis.
-
Accepts cuvettes from 10 to 40 mm pathlenght and
test tubes (o16mm)
-
Uniquely, integral cuvette tray provided convenient
storage and a rack for use during analysis,
improving work throughout and reducing spillage.
-
Large easy to read graphics display of measurement
data and scans in a variety of formats. |
|

Rapid,
easy access to sample compartment yet unaffected
by stray light |

Sample
holder is conveniently removable to enable easy
cleaning. |
|
LIGHTWAVE
SPECIFICATIONS |
|
Lightwave
UV/Vis Diode Array Spectrophotometer
Specification
S2000 Lightwave
Optical
arrangement
Single beam, diode array (512 pixel) using Rowland
Circle optics with flat field corrected concave
grating.
Wavelength
range 200 –825nm
Spectral
bandpass 5nm
Resolution/Bandwidth
5nm
Stray
Light <0.1% at 220nm NAI and 340nm NaNO2
Wavelength
Accuracy ±1.5nm
Wavelength
reproducibility Better
than ±0.6nm
Photometric
modes Abs, %T, Concn, .Abs/min,
.Abs*factor/min
Photometric
range -0.3
- 1.999A 0 - 199.9%T
Photometric
Reproducibility
±0.002Abs (0 to 0.5Abs) ±0.007Abs (0.5-1.0Abs)
Photometric
Accuracy
±0.003Abs (0 to 0.5Abs) ±0.01Abs (0.5-1.0Abs)
Slew
rate Full
spectrum measured simultaneously, measurement time
<5 sec
Zero
stability
±0.003Abs/hour after 20min warm up
Noise
±0.001Abs at 0A 500nm
Modes
of operation Scanning
Ratio (A260/A280 user adjustable) Multi Wavelength
Absorbance % Transmission Concentration (linear or
curved calibration) Kinetics Programmable - up to
99 methods stored in battery backed memory
Display
240 x 128 pixel dot matrix LCD
Output
RS232 to WPA printer for text and graphics RS232
ASCII for PC data logging
Light
sources
Pulsed Deuterium and Tungsten halogen lamps giving
life in excess of 15000 hours
Sample
compartment 40mm
pathlength. Accepts 10mm and 40mm cuvettes, 15mm
beam height. Accepts standard, semi-micro, micro
and ultra micro cuvettes downto 50/70 ハl.
Dimensions
(l x w x d) Approx. 380 x 140 x 275 mm
Weight
4.5kg
Electrical
Requirements 90-250V
50/60Hz Max 100VA |
|
|
|
|
|