Hemo - Control:
|
||||
The Hemo_Control system enables the fast and precise determination of hemoglobin and hematocrit in blood. It consists of a handy photometer, operating with microcuvettes (dry chemistry). Power supply by internal, rechargeable batteries or regular mains (100-250 V, 50/60 Hz). After the uptake of a small venous, arterial or capillary blood sample using microcuvette it has to be inserted into the Hemo_Control photometer. Within a few seconds the current level of hemoglobin and hematocrit is displayed, added by other parameters, limitations and patient information on demand. |
Technical Data - Hemo_Control:
Measurement Range |
0 – 25,6 g/dL |
Sample volume |
8 µL |
Sample material |
Venous, arterial or capillary human blood |
Linearity |
0-20 g/dL ± 0,3 g/dL; >20 g/dL ± 0,7 g/dL |
Precision |
≤ 2 % |
Measurement period |
Approx. 25 sec. |
Memory |
Up to 4,000 results including date and time, patient type |
Calibration |
In-plant according to NCCLS standard |
Quality control |
Control cuvette (included), 3 different control solutions, Hb-con low, Hb-con norm, Hb-con high |
Microcuvette |
Hemoglobin microcuvette |
Measurement principle |
Photometric, azidemethemoglobin |
Selftest |
Automatic |
Power |
Power supply, 100 - 250 VAC/50 - 60 Hz, Integrated battery (approx. 100 h in use) |
Interface |
RS 232 C (Printer), PC (EDP cable) |
Size (L x W x H) |
160 x 160 x 68 mm |
Weight |
approx. 700g |
Humidity |
Humidity of < 90 % is advisable for the use of the device |
Ambient temperature |
+15 °C to +40 °C |
Storage temperature device |
0 °C to +50 °C |
Storage temperature cuvette |
+15 °C to +30 °C |
Consumables (standard) Hemo_Control
Hemoglobin microcuvette, |
|||
SurgiLance® safety lancets orange |
|||